Artificial Respirator VG70

Make Your Best Future
We are the company in the world

ARTIFICIAL RESPIRATOR VG70 TECHNICAL DATA

Patient typeAdult and pediatric patient

Ventilation modes VCV(A/C)
  PCV(A/C)
  PRVC(optional)
  SIMV(VCV)+PSV
  SIMV(PCV)+PSV
  SIMV(PRVC)+PSV
  SPONT/CPAP+PSV
  BIVENT/APRV+PSV (optional)
  NIV/CPAP
  NIV-T
  NIV-S/T
Enhancements Apnea ventilation
  Lung mechanic
  Pressure and Flow trigger
  Automatic Tube Compensation (ATC)
  Smart suction
  Manual breath
  Insp/Exp hold
  Waveform freeze
  Nebulization
  Non-invasive ventilation (NIV) et CO2 measurement
Parameter settings:  
Ventilation frequency (f)VCV: 1-80/min (adult)
  2-80/min   (pediatric)
   
NIV Ventilation frequency (f): 4-20/min (adult)
  4-40/min (pediatric)
   
Inspiratory time (Ti): 0.2-9 s(adult)
  0.2-5 s(pediatric)
   
Tidal volume (VT): 50-2000ml  (adult)
  20-300ml(pediatric)
   
Tpause: 0-4 s(adult)
  0-2.5 s(pediatric)
Ventilation frequency (f)SIMV:1-40/min  
Inspiratory pressure (Pinsp): 5-70 cmH2O
Psupp: 0-70 cmH2O
Psupp NIV: 0 to(50-PEEP )cmH2O
PEEP: 0-35 cmH2O
PEEP in NIV: 2-20 cmH2O
CPAP in NIV: 2-20 cmH2O
Rise time (Slope) 0-2 s
O2 concentration (FiO2): 21%-100%
Trigger sensitivity: 0.5 to 20 L/min(Flow trigger)
  -20 to 0 cmH2O (Pressure trigger)
Exp. sensitivity(% of peak flow): 5%-80%
BIVENT:  
Thigh: 0.2-30 s
Tlow: 0.2-30 s
Phigh: 5-60 cmH2O
Plow: 0-35 cmH2O
Automatic Tube Compensation (ATC):  
Inner tube diameter: 5-12mm (adult)
  2.5-8mm(pediatric)
   
Tube type: Endotracheal tube et Tracheostomy tube Trach
Degree of compensation: 0-100%
Measured values displayed:  
Airway pressure measurement: Plateau pressure (Pplat)
  Positive end-expiratory pressure (PEEP)
  Peak inspiratory pressure (Ppeak)
  Mean airway pressure (Pmean)
  Min. airway pressure (Pmin)
   
Flow measurement: Expiratory minute volume (MVe)
  Spontaneous expiratory minute volume (MVespont)
  Inspiratory tidal volume (VTi)
  Expiratory tidal volume (VTe)
   
Frequency measurement: Total respiratory frequency (ftotal)
  Spontaneous respiratory frequency (fspont)
   
O2 measurement: Inspiratory O2 concentration (FiO2)
CO2 measurement: End-expiratory CO2 concentration (etCO2)
Displayed calculated values: Compliance (C)
  Resistance (R)
  leakNIV
  RSBI
  WOB
  Tispont
  I:E
  Vdaw
  PEEPi
  Tc
  Elastance
Curve displays  
Airway pressure (t) (Paw) -20 to 100 cmH2O
Flow (t) 180 to 180 L/min
Volume (t) 0 to 3000ml
CO2 (t) 0 to 80 cmH2O
   
Pressure-Volume loop  
Pressure-Flow loop  
Flow-Volume loop  
Alarms:  
Expiratory minute volume (MVe) High / Low
Airway pressure (Paw) High / Low
Expiratory tidal volume (Vte) low
PEEP High / Low
Insp.  O2  concentration  (FiO2) High  /  Low
(automatic)  
End – expiratory CO2 concentration (etCO2) High / Low
Respiratory rate (f) High
Apnea alarm time 10 to 60 seconds, Off
Inspiration duration High
Oxygen sensor failure Yes
O2 supply pressure Low
AC failure Yes
Low battery Yes
Limited battery capacity Yes
Occlusion Yes
CO2 sensor Error/Failure
Leakage out of range Yes
Fan block Yes
Performance data:  
Control principle:  
Electronically driven and electronically controlled  
Nebulization: 30min
   
Suction Oxygen enrichment: Before aspiration of sputum 3 min
  Post aspiration of sputum 2min
   
Base flow: 5-25 L/min
Max flow: 180 L/min
Leakage compensating flow: 60L/min
OPERATING DATA:  
Main power connection: 100V-240V, 50/60Hz
Power consumption:  
110-120VAC: 2A,200VAC(Ventilator only)
  10A(Ventilator plus Auxiliary Outlets)
   
220-240VAC: 1A,200VAC (Ventilator only)
  8A(Ventilator plus Auxiliary Outlets)
Gas supply: O2 gauge pressure 280-600KPa
   
Language: Chinese, English, Spanish, Polish, Turkish, Russian Other languages can be customized.
Physical Specifications:    
Dimensions(W * D * H): VG70 350mm * 460mm * 415mm
  Cart 547mm * 675mm * 950mm
     
Weight: VG70 15kg
  Cart 25kg
   
Diagonal screen size: 12″TFT color touch screen
   
Input / Output ports: 2 USB ports Real-time VGA output
  1 RJ 45 Ethernet connectors
  Nurse call
  CO2 port